| Catalog Number |
7205357 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Foreign zip code: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that during ankle arthroscopy, the handpiece was heating, it was noticed that the inner blade did not rotate.Another console was used but the issue still remained.The console was reset and the port was changed but the handpiece did not work.The procedure was successfully completed without considerable delay using a back-up device.No patient injury or other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A relationship, if any, between the subject device and the reported event could not be determined.No containment or corrective actions are recommended at this time.A review of the device history record shows there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|
| |
|