• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTO RBC AND PLASMA SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTO RBC AND PLASMA SET Back to Search Results
Catalog Number 80520
Medical Device Problem Codes Use of Device Problem (1670); Inadequate User Interface (2958)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/13/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation: a review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.The amount of citrate in as3 is 21.7mmol/l versus 113mmol/l for acda.The run data summary was analyzed for this event.Using data from the run summary, it was calculated by terumo bct that the total ac infusion rate was rate 0.80 ml/min/ltbv for this event.The replacement fluid line includes an iso 1135-4 compliant spike that is green in color to distinguish it from other fluid addition lines.When replacement fluid is enabled for a black stamp set which has a replacement line, the gui instructs the operator to connect saline to the replacement solution line and confirm the connected bag is saline.The trima accel operator's manual instructs the operator only to use normal (0.9%) saline as replacement solution, and not use saline containers larger than 1 liter.Investigation is in process.A follow-up report will be provided.
 
Event or Problem Description
The customer reported that during an auto redo blood cell (rbc) plasma donation procedure on trima, the operator attached as-3 to the saline line in error.The device did not detect the error until the trima had transfused approximately 200 ml of as-3 during saline return.The operator was alerted to the issue when the trima had a level sensor error due to the as-3 bag being empty.Full donor identifier: (b)(6).The donor was released with no side effects following the donation.The customer followed up the following day to report the donor suffered no ill effects from the donation.The customer declined to provide the donor's age.The plasma rbc collection set is not available for return because it was discarded by the customer.
 
Additional Manufacturer Narrative
This report is being filed to provide in correction: it was confirmed that the operator was retrained by the donor care manager on04/17/2019.Root cause: the root cause of the operator attaching as-3 to the saline line was a failure tofollow screen prompts to confirm the connected bag is saline.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Common Device Name
TRIMA ACCEL AUTO RBC AND PLASMA SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
MDR Report Key8488020
Report Number1722028-2019-00088
Device Sequence Number1104436
Product Code GKT
UDI-Device Identifier05020583805200
UDI-Public05020583805200
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
BK170157
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2020
Device Catalogue Number80520
Device Lot Number1807252230
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2019
Supplement Date Received by Manufacturer04/18/2019
Initial Report FDA Received Date04/05/2019
Supplement Report FDA Received Date04/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AS-3 LOT NUMBER 18BC2010B, PRODUCT # 777966000
Outcome Attributed to Adverse Event Other;
Patient Weight101
-
-