Catalog Number 72200616
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/27/2019
Event Type  malfunction  
Event Description
It was reported that during shoulder scope, device right button was stuck pressed down therefore device continuously run.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.
 
Manufacturer Narrative
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no physical damage was observed.Mdu failed functional testing with hand piece error using a known good control unit.Mdu did not run ¿on its own¿ during functional testing but did fail to function properly.The right button is stuck in the ¿on¿ position causing the hand piece error when inserted into test unit port.The complaint of running continuously was probably caused by the stuck button.The root cause of hand piece error has been determined to be a defective button switch.Factors that could have contributed to the reported event include a burr in the bushing, a bad finish on the stem, or a defective spring.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key8490020
MDR Text Key141312426
Report Number1643264-2019-00256
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public03596010555472
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 05/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2019
Initial Date Manufacturer Received 03/28/2019
Initial Date FDA Received04/07/2019
Supplement Dates Manufacturer Received05/21/2019
Supplement Dates FDA Received05/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-