SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Catalog Number 72200616 |
Device Problem
Physical Resistance/Sticking (4012)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/27/2019 |
Event Type
malfunction
|
Event Description
|
It was reported that during shoulder scope, device right button was stuck pressed down therefore device continuously run.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.
|
|
Manufacturer Narrative
|
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no physical damage was observed.Mdu failed functional testing with hand piece error using a known good control unit.Mdu did not run ¿on its own¿ during functional testing but did fail to function properly.The right button is stuck in the ¿on¿ position causing the hand piece error when inserted into test unit port.The complaint of running continuously was probably caused by the stuck button.The root cause of hand piece error has been determined to be a defective button switch.Factors that could have contributed to the reported event include a burr in the bushing, a bad finish on the stem, or a defective spring.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
|
|