| Model Number |
GELWEAVE |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Aneurysm (1708)
|
| Date of Event |
03/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).No device problem found: review of device batch history records could find no issue with batch that would lead to issue reported during procedure.Cause not established: root cause of this event could not be established from information received or review of batch history records.This is the only event of this nature reported in last 5 years.A review of information received regarding the event was carried out by clinical risk and r&d departments and conclusion given was that this event is most likely related to coagulopathy / patient issue rather than device, it is unknown if the graft not being pre-soaked in saline or rifampicin/heparin prior to procedure as recommended in ifu contributed to this event.No graft was returned and no remaining grafts from batch were available for further investigation, therefore vascutek ltd.Now considers this complaint closed however, the issue will be tracked and trended as part of the on-going complaints trending and reporting process and if an adverse trend develops action may be taken at that time.
|
| |
|
Event or Problem Description
|
|
Alleged deficiency description: clotting occurred during a procedure for ascending aortic replacement for aneurysmal disease.While cross clamp was still on around 60 minutes into the bypass clotting occurred on inner and outer surface of graft.It delayed the procedure by around 15 minutes as when clotting was noticed they had to irrigate the left ventricle and the left atrium.Patient was fully heparinised and last active clotting factor(act) was checked around 15mins prior to procedure and was 500+ seconds.No clotting was noticed on the pump filter or reservoirs.Graft not pre-soaked prior to use as recommended by ifu.
|
| |
|
Search Alerts/Recalls
|