| Model Number |
CRUOS106 |
| Medical Device Problem Codes |
Break (1069); Material Separation (1562); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.The device has been returned for evaluation; the investigation is confirmed for material separation and break of the guidewire lumen.However, the investigation is inconclusive for the reported abnormal noise issue as functional testing could not be performed due to the condition of the device.Based upon the available information, the definitive root cause is unknown.The device is labeled for single use.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction event.A review of the events indicated that model cruos106 recanalization catheter experienced an abnormal noise, material separation, and a break.This report was received from a single source.The event involved a patient with no reported injury.The patient age, weight, and sex were not provided.
|
| |
|