• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CROSSER CTO RECANALIZATION CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CROSSER CTO RECANALIZATION CATHETERS Back to Search Results
Model Number CRUOS106
Medical Device Problem Codes Break (1069); Material Separation (1562); Noise, Audible (3273)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.The device has been returned for evaluation; the investigation is confirmed for material separation and break of the guidewire lumen.However, the investigation is inconclusive for the reported abnormal noise issue as functional testing could not be performed due to the condition of the device.Based upon the available information, the definitive root cause is unknown.The device is labeled for single use.
 
Event or Problem Description
This report summarizes 1 malfunction event.A review of the events indicated that model cruos106 recanalization catheter experienced an abnormal noise, material separation, and a break.This report was received from a single source.The event involved a patient with no reported injury.The patient age, weight, and sex were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSER CTO RECANALIZATION CATHETERS
Common Device Name
RECANALIZATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8491717
Report Number2020394-2019-00252
Device Sequence Number1921979
Product Code PDU
UDI-Device Identifier10801741125611
UDI-Public(01)10801741125611
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberCRUOS106
Device Catalogue NumberCRUOS106
Device Lot NumberGFCQ1869
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2019
Initial Report FDA Received Date04/08/2019
Patient Sequence Number1
-
-