| Catalog Number |
4125127100 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/19/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A follow up report will be filed once the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that the blade broke at the mount during a total knee replacement.It was also reported that there were no delays and no adverse consequences as a result of this event.It was further reported that the broken fragments were located in the drapes and the procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results indicate that the blade break was likely due to excessive force and/ or the application of a bending moment along with metal contact while cutting.The ifu of handpieces associated with this device warn the user against this type of use: "do not apply excessive pressure, such as bending or prying, with the blade.Excessive pressure may bend or fracture the blade and result in tissue damage, loss of tactile control, and/or the ejection of blade fragments at a high velocity".The quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that the blade broke at the mount during a total knee replacement.It was also reported that there were no delays and no adverse consequences as a result of this event.It was further reported that the broken fragments were located in the drapes and the procedure was completed successfully.
|
| |
|