ST. JUDE MEDICAL - NEUROMODULATION EON MINI CHARGING SYSTEM; SCS CHARGING SYSTEM Back to Search Results
Model Number 3721
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Type  Injury  
Manufacturer Narrative
This charger model was associated with a field correction.Device information is unknown at this time.The investigation results will be provided in the final report.
 
Event Description
Device 2 of 2 : reference mfr report: 1627487-2019-03892.It was reported that the patient experienced heating sensation, but the patient could not recall if it was only during recharging the ipg.The patient received a recall letter (heating while charging), and she stopped using the device.To address the issue, the physician explanted and replaced the ipg with a new model.On 01 august 2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.An increase in prior non-reported heating while charging events and other non-reported events was expected.
 
Event Description
Device 2 of 2.Reference mfr report: 1627487-2019-03892.
 
Event Description
Device 2 of 2.Reference mfr report: 1627487-2019-03892.Follow up information identified the patient¿s burning has resolved with ipg replacement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI CHARGING SYSTEM
Type of Device
SCS CHARGING SYSTEM
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key8492345
MDR Text Key141227178
Report Number1627487-2019-04328
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 04/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/08/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2012
Device Model Number3721
Device Catalogue Number3721
Device Lot Number3279546
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/22/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/04/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-001-C
Patient Sequence Number1
Treatment
MODEL 3788, SCS IPG THERAPY
Patient Outcome(s) Other;
-
-