• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 ATTUNE REV TIB CEM CONICAL RMR; ATTUNE STEMMABLE TIBIA INSTRUMENTS : REAMERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 ATTUNE REV TIB CEM CONICAL RMR; ATTUNE STEMMABLE TIBIA INSTRUMENTS : REAMERS Back to Search Results
Catalog Number 250620103
Medical Device Problem Codes Peeled/Delaminated (1454); Mechanical Jam (2983); Scratched Material (3020)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 03/18/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device is an instrument and is not implanted/explanted.
 
Event or Problem Description
It was reported that the blade of the tibial cemented conical reamer (p/n: 250620103) was caught in the sri attune stemable tibia tower (manufactured by ksi) and it was stuck and the fixed tibia plate was come off during tka surgery on (b)(6) 2019.The surgeon re-placed it and repeated the same process.However, the blade and tower were scraped, and the metal powder was appeared.Thus, the surgeon stopped using the instruments to prevent health damage to the patient.The surgery was completed within a 30 minutes surgical delay.There was no adverse consequence to the patient.It seemed to be an issue of accuracy of the tibia tower and the conical reamer.No further information is available.
 
Additional Manufacturer Narrative
(b)(4).Evaluation of the returned device confirms deformation to the reamer.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: evaluation of the returned device confirms deformation to the reamer.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE REV TIB CEM CONICAL RMR
Common Device Name
ATTUNE STEMMABLE TIBIA INSTRUMENTS : REAMERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8492459
Report Number1818910-2019-90124
Device Sequence Number1471293
Product Code HTO
UDI-Device Identifier10603295440567
UDI-Public10603295440567
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
UNK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number250620103
Device Lot NumberNG64617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2019
Initial Date Received by Manufacturer 03/18/2019
Supplement Date Received by Manufacturer05/07/2019
05/07/2019
Initial Report FDA Received Date04/08/2019
Supplement Report FDA Received Date05/09/2019
05/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-