| Catalog Number |
5992025 |
| Medical Device Problem Codes |
Defective Device (2588); Patient-Device Incompatibility (2682); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Abscess (1690); Bacterial Infection (1735); Pain (1994); Injury (2348); Disability (2371)
|
| Date of Event |
11/22/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At this time no conclusions can be made.The patient's attorney alleges injuries and revision surgery; however, no details have been provided.The cause of the patient postoperative complications cannot be determined at this time.No lot number has been provided therefore a review of the manufacturing records is not possible at this time.Information is limited.Should additional information be provided a supplemental emdr will be submitted.Not returned.
|
| |
|
Event or Problem Description
|
|
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol ventralight st on (b)(6) 2017.It is also alleged by patient's attorney that on (b)(6) 2017, the patient had unspecified injuries related to a defect in the ventralight st, and underwent revision surgery.As reported, the patient is making a claim for an adverse patient outcome against the ventralight st.As reported, the attorney alleges general allegations for "past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient.".
|
| |
|
Event or Problem Description
|
|
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol ventralight st on (b)(6) 2017.It is also alleged by patient's attorney that on (b)(6) 2017, the patient had unspecified injuries related to a defect in the ventralight st and underwent revision surgery.As reported, the patient is making a claim for an adverse patient outcome against the ventralight st.As reported, the attorney alleges general allegations for "past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient." addendum per additional information provided: (b)(6) 2017 - patient was diagnosed with incarcerated ventral incisional hernia thereby underwent robotic assisted repair with implant of ventralight st with echo ps 2.Per op notes, ¿omental and small bowel adhesions were lysed.The hernia defect was identified with incarcerated small bowel were reduced.A ventralight st with echo ps 2 was placed in the defect and sutured circumferentially.¿ (b)(6) 2017 - patient was diagnosed with infected mesh and pain thereby underwent open repair with excision of ventralight st with echo ps 2.Per op notes, ¿abscess cavity was entered and suctioned out.The pus coming from the midline fascia from the area over the prior mesh was drained.The sutures and infected prior mesh were excised entirely.The fascia was closed with sutures.¿.
|
| |
|
Additional Manufacturer Narrative
|
|
At this time no conclusions can be made.The patient's attorney alleges injuries and revision surgery; however, no details have been provided.The cause of the patient postoperative complications cannot be determined at this time.No lot number has been provided therefore a review of the manufacturing records is not possible at this time.Information is limited.Addendum: h11: this supplemental emdr is submitted to document additional information provided and to correct the brand name and 510k number root cause is inconclusive; no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-aug-2017) is considered to be a best estimate.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|