No product has been returned for investigation nor were radiographs or ct images provided to confirm the alleged event.No patient bone quality report was available.It is unknown if the patient followed post-operative activity restrictions and/or if non-fusion may have contributed to alleged event.Labeling review: ".Patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." ".Potential risks identified with the use of this system, which may require additional surgery, include: ¿ bending, fracture or loosening of implant component(s) ¿ loss of fixation." ".Correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant.While proper selection can minimize risks, the size and shape of human bones present limitations on the size and strength of implants.Metallic and internal fixation devices cannot withstand the activity levels and/or loads equal to those placed on normal, healthy bone.These devices are not designed to withstand the unsupported stress of full weight or load bearing alone." ".These devices can break when subjected to the increased load associated with delayed union or nonunion.Internal fixation appliances are load-sharing devices that hold bony structures in alignment until healing occurs.If healing is delayed, or does not occur, the implant may eventually loosen, bend, or break.Loads on the device produced by load bearing and by the patient¿s activity level will dictate the longevity of the implant.".
|
|
On (b)(6) 2018, patient underwent a posterior lumbar inter-body fusion procedure with posterior fixation at l3/4 levels without any reported events.As per reporter, on (b)(6) 2019, a cage loosening and a screw back out was confirmed.On (b)(6) 2019, patient underwent a revision procedure where all implants were removed and replaced and construct was extended from l2/5.
|