| Catalog Number |
306601 |
| Medical Device Problem Codes |
Device Alarm System (1012); Difficult to Advance (2920)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Information (3190)
|
| Date of Event |
03/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received that while a smiths medical catheter was in use, device was difficult to thread.No patient injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: one smiths medical cadd legacy pump was returned for analysis in a good condition.During analysis, the reported problem could not be duplicated.Evidence in the log suggested that a possible low battery condition which was also supported by lec 1310.Service will perform preventive maintenance to correct the reported problem.Based on the evidence, the complaint was not confirmed, and the problem source of the reported event is unknown.
|
| |
|