• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE OMEGA 4.75MM PEEK KNOTLESS ANCHOR SYSTEM, DOUBLE-DOUBLE; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE OMEGA 4.75MM PEEK KNOTLESS ANCHOR SYSTEM, DOUBLE-DOUBLE; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 3910-500-472
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/18/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for investigation therefore the reported failure mode was not confirmed.The reported failure mode will be monitored for future reoccurrence.Alleged failure: screw became loose in joint probable root cause: design: inadequate raw material selection for screw.Design geometry too sensitive to withstand clinical impaction loads.Screw/driver interface not designed to maintain proper assembly process.Screw or driver not manufactured to specification.Incorrect material used during manufacturing.Screw and driver not assembled to specification application.Excessive force or leveraging of the screw against the bone.Inadequate assessment of bone quality, pilot hole not prepared.
 
Event or Problem Description
It was reported that the screw become loose in the joint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMEGA 4.75MM PEEK KNOTLESS ANCHOR SYSTEM, DOUBLE-DOUBLE
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key8495598
Report Number0002936485-2019-00136
Device Sequence Number942114
Product Code MBI
UDI-Device Identifier37613327464833
UDI-Public37613327464833
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K181083
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3910-500-472
Device Lot Number18353AG2
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2019
Initial Report FDA Received Date04/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age51 YR
-
-