STRYKER ENDOSCOPY-SAN JOSE OMEGA 4.75MM PEEK KNOTLESS ANCHOR SYSTEM, DOUBLE-DOUBLE; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
3910-500-472 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/18/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product was not returned for investigation therefore the reported failure mode was not confirmed.The reported failure mode will be monitored for future reoccurrence.Alleged failure: screw became loose in joint probable root cause: design: inadequate raw material selection for screw.Design geometry too sensitive to withstand clinical impaction loads.Screw/driver interface not designed to maintain proper assembly process.Screw or driver not manufactured to specification.Incorrect material used during manufacturing.Screw and driver not assembled to specification application.Excessive force or leveraging of the screw against the bone.Inadequate assessment of bone quality, pilot hole not prepared.
|
| |
|
Event or Problem Description
|
|
It was reported that the screw become loose in the joint.
|
| |
|