Event or Problem Description
It was reported that during an unknown procedure the device over-inflate, the pressure went above setting.The procedure was completed with a backup device with no delay or patient injury reported.
 
Additional Manufacturer Narrative
The device was received and shipped to the oem (original equipment manufacturer) for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed by the oem and found that the bezel, upper left corner, and both left and right bottom corners were cracked.The bulkhead seal to hose connection was damaged.A functional evaluation was performed by the original equipment manufacturer.The unit was tested for proper operation and performed as expected.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.An oem review of the device history records show there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.The user can reference the operations/service manual, for information regarding reducing the likelihood of potential over-pressure situations which includes the use of the smith & nephew 500 insufflator tap¿ feature and the associated tap¿ tubing set as often as possible.Do not use any other tubing set for this purpose as it may directly affect the pressure monitoring results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMITH & NEPHEW 500 INSUFFLATOR
Common Device Name
INSUFFLATOR, LAPAROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key8496266
Report Number1643264-2019-00262
Device Sequence Number14049409
Product Code HIF
UDI-Device Identifier03596010673497
UDI-Public03596010673497
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K120151
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72203994
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2019
Initial Date Received by Manufacturer 03/21/2019
Supplement Date Received by Manufacturer05/30/2019
07/17/2019
Initial Report FDA Received Date04/09/2019
Supplement Report FDA Received Date06/03/2019
07/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-