| Model Number |
TMICL12.6 |
| Medical Device Problem Codes |
Failure to Unfold or Unwrap (1669); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
03/11/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Work order search: no similar complaint type events were reported for units within the same lot.Device evaluation: lens was returned stuck in cartridge.The cartridge was returned in lens box.There was residue on product.Visual inspection found no visible damage to the device and residue on cartridge.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 12.6mm, tmicl12.6, -13.5/2.0/093 (sphere/cylinder/axis), implantable collamer lens would not open/unfold properly during injection into the right eye (od).The incision was widened and sutures were used.The lens was implanted, removed and exchange within the same surgery and the problem was resolved.The reporter stated "no problems" for patients current condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Device code 2993 in previous mdr is not applicable.(b)(4).
|
| |
|
Search Alerts/Recalls
|