CYBERONICS - HOUSTON PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problem Premature End-of-Life Indicator (1480)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/11/2015
Event Type  malfunction  
Event Description
Product analysis was completed and approved on the returned generator.A comprehensive automated electrical evaluation showed that the device performed according to functional.After the trimmed edge of the circuit board (pcba) was cleaned, a postburn electrical test was performed.Results of the postburn electrical test were magnet detection normal and magnet response test failed.Based on the postburn electrical test results, the contamination that was observed on the trimmed edge of the pcba suggest probable electrical paths (resistive path) were established between the copper edges on the trimmed edge of the pcba, which contributed to the supply current conditions.Data was reviewed that showed that the battery is depleting faster than expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
rachel kohn
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key8499478
MDR Text Key141464035
Report Number1644487-2019-00690
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750061
UDI-Public05425025750061
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/05/2017
Device Model Number106
Device Lot Number203469
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2018
Event Location Other
Date Manufacturer Received03/18/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age15 YR
-
-