Other, other text: additional information provided h3, h6 and h10.This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A device history record (dhr) review was performed and there were no relevant findings.The device was returned for analysis and has been evaluated.The returned device consisted of ten samples which were received in used condition without its original packaging.Visual and functional testing were performed.The ten samples were visually inspected individually at a distance of 12 inches to 24 inches and normal conditions of illumination.During visual inspection, no discrepancies were detected.During functional test, a syringe was used to infuse water into the ay bag.The ten samples presented leakage located in the top edge of the corner.The root cause of the reported event was due to manufacturing.During the assembly process, the ay bag was not handled properly during loading operation.Actions were taken to mitigate the reported issue: no actions were taken for the primary concern reported; however, training to production and quality personnel will be required.
|