| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) internal report # (b)(4).Returned meter evaluated with no defect found.Incorrect test strips returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage (kitchen).Test strip udi#: (b)(4).Note: manufacturer contacted customer in a follow-up call to ensure that the replacement products resolved the initial concern - able to establish contact with customer who indicated that is comfortable with results from the replacement meter.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 249, 301 and 315 mg/dl.The customer's expected fasting blood glucose test result range is 100 - 121 mg/dl.Customer stated the high readings started several weeks ago.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the kitchen.The test strip lot manufacturer's expiration date is 03/26/2020 and test strips were opened one week ago.The meter memory was reviewed for previous test result history (date/time not set): (b)(6).
|
| |
|