| Model Number |
U4200310RX |
| Medical Device Problem Codes |
Break (1069); Difficult to Insert (1316); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
For the one reported event, the one device was returned to bd and evaluated.The company is still investigating the issue at this time.
|
| |
|
Event or Problem Description
|
|
This report summarizes one malfunction event.A review of the event indicated that model u4200310rx pta balloon dilatation catheter allegedly had difficulty inserting through the sheath.These report was received from one source.Of the one event, did involved patient with no reported patient injury.The patients' age ranged from (b)(6) years, weight was (b)(6) lbs.Of the reported patient, one was female.
|
| |
|
Event or Problem Description
|
|
This report summarizes one malfunction.A review of the reported information indicated that model u4200310rx pta balloon dilatation catheter allegedly experienced a break, difficult to insert, and a detachment of device component.This information was received from a single source.The malfunction involved a patient with no reported consequence.The patient was reported as a 74-year-old female weighing 120 lbs.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot number for the malfunction was provided and a lot history review was performed.The device has been returned for evaluation; the investigation is confirmed for detachment.It is likely the user perceived the detachment as a break.The investigation is inconclusive for the alleged insertion issue as functional testing could not be conducted.Based upon the available information, the definitive root cause is unknown.The device is labeled for single use.
|
| |
|
Search Alerts/Recalls
|