| Catalog Number |
LX105 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Only event year known: 2019.Possible serial numbers: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the jaws of the clip appliers are misaligned, resulting in the clips falling out of the jaws of the instrument during use or not locking on tissue during use.No patient involvement was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Device analysis: the analysis results confirmed that the lx105 device was returned nonfunctional as the jaws were misaligned; therefore, the clips could not be loaded into the jaws.It is possible that the damaged was due to an improper handling of the device.Please note that as stated in the ifu: "all surgical instruments are subject to a degree of wear and tear because of normal use.A regular and precise visual check of the instrument should be made before each use.The batch history records were reviewed and certed by external manufacturing that the manufacturing criteria was met prior to the release of the equipment. the certificate records are accessible through external manufacturing. .
|
| |
|