| Model Number |
E SERIES |
| Medical Device Problem Codes |
Pacing Problem (1439); Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional procode=dro.The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a loose flex cable to the connector on the system board.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed "defib fault 72" "defib disabled" and "pacer disabled" messages.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|