• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIATRS45AXT
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during coil video-assisted thoracoscopic surgery (vats) wedge resection with bronchoscopy, there were two(2) staplers and four(4) reloads that did not fire correctly or cut through tissue.Another device was used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the cartridge noted that the reload was partially fired and the anvil clamping mechanisms was deformed.Functional testing of the reload found no abnormalities.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the deformed anvil clamping mechanism may occur in any of the following circumstances: firing over tissue that is beyond the recommended thickness range.Firing with an obstacle incorporated in the jaws.In either of these circumstances, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly, and tissue may not be fully transected.The information booklet which accompanies each product shipment offers the following as a warning and recaution." preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range thickness for the staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in corresponding selection of staple size.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key8504665
Report Number1219930-2019-02243
Device Sequence Number16198756
Product Code GDW
UDI-Device Identifier10884521190993
UDI-Public10884521190993
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K133938
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2021
Device Model NumberEGIATRS45AXT
Device Catalogue NumberEGIATRS45AXT
Device Lot NumberN8D1066KX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2019
Initial Date Received by Manufacturer 03/20/2019
Supplement Date Received by Manufacturer05/30/2019
Initial Report FDA Received Date04/11/2019
Supplement Report FDA Received Date06/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-