MAUDE Adverse Event Report: ENCORE MEDICAL L.P. PS KNEE; INSERT, POSTERIOR STABILIZED, SERIES 500, SIZE8 13MM CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. PS KNEE; INSERT, POSTERIOR STABILIZED, SERIES 500, SIZE8 13MM CM Back to Search Results
Model Number 360-13-508
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/14/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was due to the screw backing out of the insert.The previous surgery and the surgery detailed in this investigation occurred 5.5 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to the screw backing out of the insert.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - due to the screw backing out of the insert.The surgeon removed and replaced insert with a djo product.The screw and patella were removed and replaced by another vendor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS KNEE
Common Device Name
INSERT, POSTERIOR STABILIZED, SERIES 500, SIZE8 13MM CM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8504757
Report Number1644408-2019-00287
Device Sequence Number1611912
Product Code JWH
UDI-Device Identifier00888912116428
UDI-Public(01)00888912116428
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K933539
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/28/2014
Device Model Number360-13-508
Device Catalogue Number360-13-508
Device Lot Number53990764
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2019
Initial Report FDA Received Date04/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
120-01-032, LOT 469A1033
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
-
-