| Model Number |
D013 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/13/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
| |
|
Event or Problem Description
|
|
It was reported that a red alert triggered on this patient's communicator, indicating a change on this implantable cardioverter defibrillator (icd).No further information was able to be obtained, but at this time, the device remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
This report is being filed to provide product investigation results.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed, and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Search Alerts/Recalls
|