• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VICRYL SUTURE 27"(70CM) 0 VIO; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VICRYL SUTURE 27"(70CM) 0 VIO; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number J329H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that the patient underwent an unknown procedure on an unknown date and suture was used.The needle broke in half when closing the incision.The broken piece was found, and the procedure was not delayed.Another like device was used to complete the procedure.There were no adverse patient consequences reported.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary: it was reported performance needle breakage.A failed suture needle was submitted for fractographic evaluation.A fracture was observed in the body of the needle.The needle was received in two pieces, only one of the mating fracture surfaces was examined for this evaluation.A microscope was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations in order to determine the fracture mode.The evaluation revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile overload failure.Mechanical damage and a cup-and-cone shaped fracture observed coincidental to the fracture provide additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.The manufacturing records couldn't be reviewed as the batch number is unknown.The evidence of this examination indicates that the breakage occurred from mechanical deformation, indents and scratches, due to gripping, bending and overstressing of the needle. there is no evidence of any material flaw that would cause premature failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VICRYL SUTURE 27"(70CM) 0 VIO
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key8506082
Report Number2210968-2019-80286
Device Sequence Number1628669
Product Code GAM
UDI-Device Identifier10705031038875
UDI-Public10705031038875
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberJ329H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2019
Initial Date Received by Manufacturer 03/20/2019
Supplement Date Received by Manufacturer04/16/2019
Initial Report FDA Received Date04/11/2019
Supplement Report FDA Received Date05/13/2019
Patient Sequence Number1
-
-