• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #4 L-CEM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #4 L-CEM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5510-F-401
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Codes Injury (2348); Ambulation or Postural Difficulties (2544)
Date of Event 11/04/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The information in this report was provided by stryker orthopaedics legal affairs department.No additional information is available at this time due to the ongoing litigation.Should additional information become available, the evaluation summary will be submitted in a supplemental report.Product not returned.
 
Event or Problem Description
It was reported through the communication of an attorney that allegedly the patient was revised due to "instability of the left knee".
 
Additional Manufacturer Narrative
The information in this report was provided by stryker orthopaedics legal affairs department.An event regarding instability involving a triathlon femoral component was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the reported event nor the exact cause could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.Product surveillance will continue to monitor for trends.Corrective action/preventive action: no action is required at this time as there is no indication to suggest a product nonconformity or unanticipated hazard.
 
Event or Problem Description
It was reported through the communication of an attorney that allegedly the patient was revised due to "instability of the left knee.".
 
Additional Manufacturer Narrative
Reported event an event regarding instability involving a triathlon femoral component was reported.The event was confirmed by medical review.Method & results -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-medical records received and evaluation: a review of the provided medical records by a clinical consultant indicated: i can confirm that the event occurred since i was able to review the operation reports and i reviewed pre op and post op x-rays.Regarding the possible root cause of this event, i cannot determine this for certain.I don¿t see any abnormality with the implant itself.The development of postoperative instability is a well-known complication.Operative technique, soft tissue balancing as well as activity level and body habitus are key in the longevity of a total knee replacement.-product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
It was reported through the communication of an attorney that allegedly the patient was revised due to "instability of the left knee".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR FEM COMP #4 L-CEM
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8506483
Report Number0002249697-2019-01593
Device Sequence Number18887773
Product Code MBH
UDI-Device Identifier07613327040272
UDI-Public07613327040272
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2016
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2017
Device Model Number5510-F-401
Device Catalogue Number5510F401
Device Lot NumberS8R3K
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/15/2019
Supplement Date Received by Manufacturer05/22/2019
08/30/2021
Initial Report FDA Received Date04/11/2019
Supplement Report FDA Received Date06/17/2019
09/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age65 YR
-
-