Model Number 5551-G-320
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Codes Injury (2348); Ambulation or Postural Difficulties (2544)
Date of Event 11/04/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.The information in this report was provided by stryker orthopaedics legal affairs department.No additional information is available at this time due to the ongoing litigation.Should additional information become available, the evaluation summary will be submitted in a supplemental report.Product not returned.
 
Event or Problem Description
It was reported through the communication of an attorney that allegedly the patient was revised due to "instability of the left knee".
 
Event or Problem Description
It was reported through the communication of an attorney that allegedly the patient was revised due to "instability of the left knee.".
 
Additional Manufacturer Narrative
Reported event: an event regarding instability involving a triathlon patella was reported.The event was confirmed by medical review.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Medical records received and evaluation: a review of the provided medical records by a clinical consultant indicated: i can confirm that the event occurred since i was able to review the operation reports and i reviewed pre op and post op x-rays.Regarding the possible root cause of this event, i cannot determine this for certain.I don¿t see any abnormality with the implant itself.The development of postoperative instability is a well-known complication.Operative technique, soft tissue balancing as well as activity level and body habitus are key in the longevity of a total knee replacement.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON ASYMMETRIC X3 PATELLA
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8506486
Report Number0002249697-2019-01592
Device Sequence Number1571432
Product Code MBH
UDI-Device Identifier07613327014815
UDI-Public07613327014815
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2016
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2017
Device Model Number5551-G-320
Device Catalogue Number5551-G-320
Device Lot NumberY99X
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/15/2019
Supplement Date Received by Manufacturer08/30/2021
Initial Report FDA Received Date04/11/2019
Supplement Report FDA Received Date09/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age65 YR
-
-