• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER GMBH BIOLOX DELTA, CERAMIC FEMORAL HEAD, L, ø 32/+3.5, TAPER 12/14; BIOLOX DELTA CERAMIC FEMORAL HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER GMBH BIOLOX DELTA, CERAMIC FEMORAL HEAD, L, ø 32/+3.5, TAPER 12/14; BIOLOX DELTA CERAMIC FEMORAL HEAD Back to Search Results
Model Number N/A
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/18/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.The manufacturer did not receive x-rays, or other source documents for review.Device history record (dhr) was reviewed and no discrepancies were found a cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available an updated report will be submitted.(b)(4).Discarded by hospital.
 
Event or Problem Description
It was reported that the patient underwent a revision surgery due to implant loosening.Attempts to obtain additional information have been made; however, no more is available.
 
Event or Problem Description
Please refer to report 0009613350-2019-00213.
 
Additional Manufacturer Narrative
This follow-up report is being filled to relay investigation result.Dhr review: the device manufacturing quality records indicate that the released components met all requirements to perform as intended.Trend analysis: no trigger considering the following event is identified: subluxation.Event description: it was reported that patient underwent an initial left thr surgery on (b)(6) 2013 and revised on (b)(6) 2015 in which the biolox delta head was implanted.Finally, it was explanted on (b)(6) 2019 due to loosening.Surgeon stated that the tension on the ceramic head and the socket was loose, not devices.Review of received data: no medical data such as x-rays, surgical notes or any other case-relevant documents received.Device analysis: no product was returned to zimmer biomet for in-depth analysis.Review of product documentation: this device is intended for treatment.Compatibility: compatibility check could not be performed as only one product had been reported.Surgical technique is not reviewed as no surgical report was received to compare whether the surgical technique was followed or not.Conclusion: review of the device history records for the product did not identify any deviations or anomalies related to the reported event.The investigation results did not identify a non-conformance or a complaint out of box (coob).Neither x-rays, operative notes, office visit notes, nor devices or photos of the explanted implant was received.Patient factors that may affect the performance of the components such as bone quality, activity level, type of activity (low impact vs.High impact), soft tissue laxity and relevant medical history are unknown.Adherence to rehabilitation protocol is unknown.Other components of the thr system are also unknown.Therefore, in the absence of required medical data the complaint could not be confirmed.In conclusion, due to significant lack of information, it is impossible to perform a meaningful analysis of the reported event.Therefore, an exact root cause cannot be determined.However, the most likely reason is increased soft tissue laxity since the doctor stated that the head/cup connection is loose and not the devices.The need for corrective measures is not indicated and zimmer gmbh considers this case as closed.Zimmer biomet¿s reference number of this file is (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOLOX DELTA, CERAMIC FEMORAL HEAD, L, ø 32/+3.5, TAPER 12/14
Common Device Name
BIOLOX DELTA CERAMIC FEMORAL HEAD
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
MDR Report Key8507623
Report Number0009613350-2019-00213
Device Sequence Number11017692
Product Code LZO
UDI-Device Identifier00889024430334
UDI-Public00889024430334
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K071535
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2023
Device Model NumberN/A
Device Catalogue Number00-8775-032-03
Device Lot Number2728636
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/18/2019
Supplement Date Received by Manufacturer10/04/2019
Initial Report FDA Received Date04/12/2019
Supplement Report FDA Received Date10/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age41 YR
-
-