LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-02-15
Device Problem Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/11/2019
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Livanova deutschland manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a defective encoder.The technician replaced the encoder to resolve the reported issue.The defective encoder was returned to livanova deutschland for further investigation.During the evaluation the reported issue could be confirmed but as the device is not manufactured by livanova a specific root cause could not be determined.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.
 
Event Description
Livanova deutschland received a report that a knob of a centrifugal pump system with tubing clamp was hard to be rotated during procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
njemile crawley
14401 w. 65th way
arvada, CO 80004
2812287575
MDR Report Key8508401
MDR Text Key141751036
Report Number9611109-2019-00249
Device Sequence Number1
Product Code DWA
UDI-Device Identifier04003381790098
UDI-Public01040033817900986111808012160S14748
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 04/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-15
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/11/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-