|
Analysis found that visually, there was no damage or anomalies in the construction of the device that would have resulted in the reported event.The biological contaminants were cleaned off the emg contacts prior to electrical testing.Electrically, the resistance of the tube electrodes from end to end shall be less than 200 ohms and the actual measurements were; red 18 ohms, red [w/white band] 26 ohms, blue 22 ohms, and blue [w/white band] 15 ohms.There were no shorts between circuits or intermittent behavior while manipulating the circuits.When viewed under magnification, there were no cracks in the electrode contacts.If information is provided in the future, a supplemental report will be issued.
|