Model Number M00565040
Medical Device Problem Codes Difficult or Delayed Positioning (1157); Premature Activation (1484)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/20/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age or date of birth: patient's exact age is unknown; however it was reported that the patient was over the age of 18.(b)(4).The complainant indicated that the suspect device remains implanted and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2019 that a wallflex colonic stent was implanted in the sigmoid colon to treat a 6-7 cm mass during a colonoscopy with stent placement procedure performed on (b)(6) 2019.Reportedly, the patient's anatomy was tortuous and was not dilated prior to stent placement.According to the complainant, during the procedure, it was difficult to unsheathe the stent.The stent suddenly deployed through the stricture and moved to the proximal part of the colon.A longer wallflex colonic stent was placed stent-in-stent within the initial stent to complete the procedure.There were no patient complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good with passing gas.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX COLONIC
Common Device Name
STENT, COLONIC, METALIC, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8509485
Report Number3005099803-2019-01745
Device Sequence Number16457167
Product Code MQR
UDI-Device Identifier08714729456513
UDI-Public08714729456513
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K061877
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Model NumberM00565040
Device Catalogue Number6504
Device Lot Number0023279444
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2019
Initial Report FDA Received Date04/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-