CODMAN AND SHURTLEFF, INC OG CMPLX XTRASOFT COIL 2X2; NEUROVASCULAR EMBOLIZATION DEVICE
|
Back to Search Results |
|
| Catalog Number |
640CX0202 |
| Medical Device Problem Code |
Stretched (1601)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Information regarding patient age/date of birth, gender, weight, height, medical history, race, and ethnicity was not provided.Procode: krd/hcg.Initial reporter: the customer contact information, including phone, fax, and e-mail address, was not reported.The product is available for evaluation and testing; however, it has not been received to date (which is indicated as "other").A review of the manufacturing documentation associated with this lot presented no issues during the manufacturing or inspection process that can be related to the reported complaint.No non-conformances related to the reported complaint condition were identified.Additional information will be submitted within 30 days of receipt.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|
Event or Problem Description
|
|
As reported by a healthcare professional, during a coil embolization of a 4.1mm x 4mm right anterior communicating artery aneurysm, the physician found it difficult to advance the 2mm x 2cm orbit galaxy xtrasoft (640cx0202/30096541) coil through the body/shaft (approx 10cm) of the unspecified microcatheter and noticed the embolic coil had stretched under x-ray.The coil was removed from the microcatheter, and the procedure was completed with no patient consequence or flow restriction/reduction.The event did not result in a loss of cerebral target position.The procedural delay was not considered clinically significant.The device was stored and handled according to the instructions for use (ifu).The procedure was conducted in accordance with the ifu and an adequate and continuous flush was maintained through the microcatheter.There were no kinks in the concomitant microcatheter that may have contributed to the event.The coil had not exited or deployed from the microcatheter.The embolic coil was still attached to the coil delivery system when removed from the patient.The device will be returned for evaluation.No further information could be obtained.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).The purpose of this mdr submission is to report that the product has been returned for evaluation and testing; however, the engineering evaluation has not been completed.Additional information will be submitted within 30 days of receipt.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).[complaint conclusion] as reported by a healthcare professional, during a coil embolization of a 4.1mm x 4mm right anterior communicating artery aneurysm, the physician found it difficult to advance the 2mm x 2cm orbit galaxy xtrasoft (640cx0202/30096541) coil through the body/shaft (approx 10cm) of the unspecified microcatheter and noticed the embolic coil had stretched under x-ray.The coil was removed from the microcatheter, and the procedure was completed with no patient consequence or flow restriction/reduction.The event did not result in a loss of cerebral target position.The procedural delay was not considered clinically significant.The device was stored and handled according to the instructions for use (ifu).The procedure was conducted in accordance with the ifu and an adequate and continuous flush was maintained through the microcatheter.There were no kinks in the concomitant microcatheter thast may have contributed to the event.The coil had not exited or deployed from the microcatheter and the coil was still attached to the coil delivery system when removed from the patient.No further information could be obtained.A non-sterile 2mm x 2cm orbit galaxy xtrasoft was received coiled inside of a pouch.Visual inspection was performed on the device.The hub of the delivery tube was found undamaged.The hypotube was found kinked at 17.5cm, 54cm, and 79cm from its proximal end.The introducer was found zipped and undamaged.The device was then examined under a microscope.The support coil and gripper were seen undamaged inside of the introducer.The embolic coil was seen stretched inside of the introducer.Advancement through a lab sample microcatheter could not be tested since the hypotube was received kinked.A manufacturing record evaluation was performed for the finished device is 30096541 number, and no non-conformances related to the reported complaint condition were identified.The customer report of resistance/friction during advancement could not be evaluated due to the kinked condition of the hypotube.The customer report of coil unraveled stretched was confirmed during microscopic inspection.The exact circumstances surrounding the observed damage to the hypotube and embolic coil cannot be determined based on the condition of the returned device; however, kinking of the hypotube and stretching of the embolic coil are indicative of the application of excessive force, possibly in an attempt to overcome the reported resistance.100% of devices are inspected for damage at final inspection.Thus, it is very unlikely that the device left the manufacturing facility with the observed damage.Neither the analysis nor the mre suggests that the failure reported could be related to the manufacturing process.Functional testing could not be performed to determine potential root causes of the event due to the condition of the returned device.In addition, without the return of the concomitant microcatheter, the cause of the reported event cannot be determined.Resistance/friction and coil unraveling/stretching are known potential product issues associated with the use of the device.The ifu contains several cautions relating to this situation, including instructions for troubleshooting the situation should it be encountered during use.Assignment of root cause for the event remains speculative and inconclusive, based on the information provided and the evidence presented by the returned device; however, it is possible that circumstances of the procedure may have contributed to the reported failure and damages noted to the returned system.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective/preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|