| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint (b)(4).Concomitant medical products: biomet microfixation sternalock blade catalog #: 73-1194, lot #: 251810; biomet microfixation sternalock pd blade catalog #: 73-1191, lot #: 263760.Therapy date - unknown.Occupation - sales representative.Initial reporter also sent report to fda: the user facility is foreign; therefore a facility medwatch report will not be available.Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2019-00218 & 0001032347-2019-00219.
|
| |
|
Event or Problem Description
|
|
It was reported by the sales rep that the small ball on top of the blade broke during a procedure.Attempts have been made and no further information has been provided.
|
| |
|
Event or Problem Description
|
|
This follow-up report is being submitted to relay additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.The blades were returned for investigation.The short blade still had the tip intact, however, the blade was clearly bent.The complaint is confirmed.Dhr was reviewed and no discrepancies were found.Investigation results concluded that the reported event was due excessive force being applied, possibly during an off-axis attempt to insert a screw.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2019-00218-1 & 0001032347-2019-00219-1.
|
| |
|
Search Alerts/Recalls
|