• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ORIGEN MINI ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ORIGEN MINI ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number D002
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/13/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) was behaving in a manner consistent with a recorded code 1003 indicative of premature battery depletion and triggered elective replacement indicator (eri) pre-implant.No patient involvement.At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORIGEN MINI ICD DR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key8512380
Report Number2124215-2019-06162
Device Sequence Number16556866
Product Code LWS
UDI-Device Identifier00802526533747
UDI-Public00802526533747
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P960040/S306
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/04/2021
Device Model NumberD002
Device Catalogue NumberD002
Device Lot Number118811
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2019
Initial Report FDA Received Date04/13/2019
Patient Sequence Number1
-
-