MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. CYSTO-NEPHRO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-VH
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/21/2019
Event Type  Malfunction  
Manufacturer Narrative

The subject device in this report has not been returned to omsc for evaluation. Omsc reviewed the manufacture history (dhr) of the device and confirmed no irregularity. The exact cause of the reported event could not be conclusively determined at this time. If additional information becomes available, this report will be supplemented.

 
Event Description

Olympus medical systems corp. (omsc) was informed that as a result of routine multiple microbiological testing by the user facility, following microbes were detected from the sample collected from the subject device. [first time; (b)(6) 2018] - all channels: gram-positive bacteria (2cfu/endoscope) and staphylococcus coagulase negative (3cfu/endoscope), [second time; (b)(6) 2018] - all channels: gram-positive bacteria (1cfu/endoscope), [third time; (b)(6) 2019] - all channels: staphylococcus coagulase negative (1cfu/endoscope), [forth time; (b)(6) 2018] - all channels: gram-positive bacteria (31cfu/endoscope). The device had been manually reprocessed using peracetic acid. There was no report of infection associated with this report.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCYSTO-NEPHRO VIDEOSCOPE
Type of DeviceCYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key8512687
MDR Text Key143207237
Report Number8010047-2019-01678
Device Sequence Number1
Product Code FAJ
Combination Product (Y/N)N
Reporter Country CodeFR
PMA/PMN NumberK062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type FOREIGN,USER FACILITY
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 05/08/2019
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received04/14/2019
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberCYF-VH
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received04/19/2019
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured03/19/2018
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-