| Model Number |
AR40E |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
03/19/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If explanted, give date: not applicable, remains implanted in the eye.(b)(4).Device evaluation: the product testing could not be performed as the product was not returned as the lens remains implanted.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the review.The product was manufactured and released according to specifications.The search revealed that no additional complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that initially there was loading error with sensar monofocal intraocular lens, so surgeon thought of using sensar monofocal intraocular lens with same model and same diopter (model ar40e +25.0 diopter).However, this lens had broken haptic during surgery.So, surgeon implanted initial lens successfully using forceps by hand and no cartridge was used.During the same procedure an incision enlargement and sutures were required.No further information provided.This report will capture information for lens that remains implanted.A separate report is being submitted for lens that had a broken haptic issue.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|
Search Alerts/Recalls
|