Model Number AR40E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 03/19/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If explanted, give date: not applicable, remains implanted in the eye.(b)(4).Device evaluation: the product testing could not be performed as the product was not returned as the lens remains implanted.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the review.The product was manufactured and released according to specifications.The search revealed that no additional complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision has been submitted.
 
Event or Problem Description
It was reported that initially there was loading error with sensar monofocal intraocular lens, so surgeon thought of using sensar monofocal intraocular lens with same model and same diopter (model ar40e +25.0 diopter).However, this lens had broken haptic during surgery.So, surgeon implanted initial lens successfully using forceps by hand and no cartridge was used.During the same procedure an incision enlargement and sutures were required.No further information provided.This report will capture information for lens that remains implanted.A separate report is being submitted for lens that had a broken haptic issue.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key8512879
Report Number2648035-2019-00457
Device Sequence Number13847033
Product Code HQL
UDI-Device Identifier05050474502208
UDI-Public(01)05050474502208(17)210412
Combination Product (Y/N)Y
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/11/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/12/2021
Device Model NumberAR40E
Device Catalogue NumberAR40E00250
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 03/20/2019
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date04/15/2019
Supplement Report FDA Received Date11/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-