• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US SUPER QA+ #2 ETH CP-2 *EA; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US SUPER QA+ #2 ETH CP-2 *EA; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Catalog Number 212032
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Code Available (3191)
Date of Event 03/21/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Event or Problem Description
It was reported by the affiliate via email that during an unknown procedure the 2 wires were stripped and then broke without excessive traction.A second anchor was used in a limited space because it was impossible to remove the broken anchor.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device is not being returned, multiple attempts for product return were made with no response, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.No nonconformances were identified for this part-lot number combination per qlik query executed 05/28/2019.Multiple attempts to obtain additional information were made but no response was received.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8514924
Report Number1221934-2019-56848
Device Sequence Number19323442
Product Code JDR
UDI-Device Identifier10886705001804
UDI-Public10886705001804
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K930893
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Catalogue Number212032
Device Lot Number2L06374
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/22/2019
Supplement Date Received by Manufacturer05/28/2019
Initial Report FDA Received Date04/15/2019
Supplement Report FDA Received Date06/03/2019
Patient Sequence Number1
-
-