Model Number 7311 |
Device Problem
Device Markings/Labelling Problem (2911)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/07/2019 |
Event Type
Injury
|
Event Description
|
It was reported that a labeling issue occurred.During a percutaneous coronary intervention (pci), a 5.50mm x 8mm nc emerge was used to dilate the target lesion.The device appeared to be bigger than the printed label.The procedure was completed with this device.No patient complications were reported in relation to this event and the patient status was stable.
|
|
Event Description
|
It was reported that a labeling issue occurred.During a percutaneous coronary intervention (pci), a 5.50mm x 8mm nc emerge was used to dilate the target lesion.The device appeared to be bigger than the printed label.The procedure was completed with this device.No patient complications were reported in relation to this event and the patient status was stable.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: the returned product consisted of a nc emerge balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.The device returned with the shelf box and pouch.Visual examination revealed multiple kinks along the hypotube.Microscopic examination revealed damage to the tip.There is contrast present in the inflation lumen and balloon.There is blood in the hub and guidewire lumen.The balloon was loosely folded prior to inflating for dimensional testing.Per product specification, the balloon diameter at nominal pressure of 12atm should measure between 5.17mm minimum and 5.83mm max.The balloon measured at 5.56mm.The body length of the balloon at rated burst pressure of 18atm should measure 8+/-1mm.Product analysis did not confirm any discrepancy with the size of the balloon.
|
|
Search Alerts/Recalls
|