| Catalog Number |
VR416 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/02/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A manufacturing record evaluation was performed for the finished device vr416, lj9dkhe0 number, and no non-conformances related to the reported complaint condition were identified.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an unknown procedure on (b)(6) 2019 and suture was used.During the procedure, the suture was broken when tying.There were no adverse patient consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).It was performance breakage suture.An unopened sample of product code vr416, lot lj9dkhe0 were returned for analysis.During the visual inspection of the sample, no defects were found on the package.The sample was opened, and the swage and attachment area were noted to be as expected.The suture was dispensed without problem and examined along of the strand and no defects, damaged or suture breakage were noted.A functional test was performed and the tensile strength force was above the minimum requirement.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the representative sample, no performance - breakage suture was found and the tested sample met the finished goods requirements.
|
| |
|