| Model Number |
24674 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in a vessel in the arm.A 9.0 x 80, 75cm mustang balloon catheter was advanced for dilation.However, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient's status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.Device evaluated by mfr.: the device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A microscopic examination identified a longitudinal tear beginning approximately 9mm distal of the distal markerband and extending approximately 48mm proximally across the balloon material.An examination of the balloon material and markerband identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination observed no issues with the tip of the device which could potentially have contributed to the complaint incident.A visual and tactile examination found no kinks or damage along the shaft of the device.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in a vessel in the arm.A 9.0 x 80, 75cm mustang balloon catheter was advanced for dilation.However, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient's status was stable.
|
| |
|