| Model Number |
HT70 |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/20/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the service engineer evaluated the device and reconnected the alarm cable to the single board computer assembly.After reconnecting the cable, the ventilator passed all testing per manufacturing specification and was placed back into clinical use.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that, during use on patient, the ht70 ventilator did not generated an audio alarm when the patient circuit was disconnected to use the nebulizer.The patient was removed from the ventilator and placed on an alternate ventilator with no harm reported.
|
| |
|