COVIDIEN HT70 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number HT70
Medical Device Problem Code Defective Alarm (1014)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/20/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation summary: the service engineer evaluated the device and reconnected the alarm cable to the single board computer assembly.After reconnecting the cable, the ventilator passed all testing per manufacturing specification and was placed back into clinical use.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that, during use on patient, the ht70 ventilator did not generated an audio alarm when the patient circuit was disconnected to use the nebulizer.The patient was removed from the ventilator and placed on an alternate ventilator with no harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HT70 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
michael collins rd mervue
galway,ie
MDR Report Key8515986
Report Number8020893-2019-00086
Device Sequence Number2798142
Product Code CBK
UDI-Device Identifier10884521542266
UDI-Public10884521542266
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K111146
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHT70
Device Catalogue NumberHT70PM-WW-NA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2019
Initial Report FDA Received Date04/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age80 YR
Patient Weight81
-
-