BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 2.5/9; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 401730
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/21/2019
Type of Reportable Event Malfunction
Event or Problem Description
The orsiro drug-eluting stent was positioned in the calcified and torturous medial cx.After balloon inflation it was not clear whether the orsiro stent was deployed.Therefore the device was withdrawn and outside of the patient it was noticed that the stent had slipped back to the proximal shaft of the catheter.
 
Additional Manufacturer Narrative
The device was not returned to biotronik.Therefore no technical investigation on the subject could be performed.The manufacturing history was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.The device in question was manufactured according to specifications and successfully passed all in process controls as well as the final inspection.Based on the conducted investigations of the device being subject to this complaint, no manufacturing related root cause could be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 2.5/9
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key8516667
Report Number1028232-2019-01491
Device Sequence Number1651516
Product Code NIQ
UDI-Device Identifier07640130438859
UDI-Public07640130438859
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number401730
Device Catalogue NumberSEE MODEL NO.
Device Lot Number09184657
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/12/2019
Supplement Date Received by Manufacturer05/20/2019
Initial Report FDA Received Date04/15/2019
Supplement Report FDA Received Date05/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-