• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAH; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAH; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAH
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 04/01/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer called and reported that they received emergency medical assistance due to low blood glucose on (b)(6) 2019 with blood glucose of 33 mg/dl at the time of the incident.The customer was at 220 mg/dl at the time of the call.The customer used food, glucose tablets, and was given intravenous glucose to treat.The customer experienced symptoms such as fainting.The customer was wearing the insulin pump during the incident.The customer believes the pump was over delivering.Troubleshooting was completed but did not resolve the issue.The customer had a low of under 20 mg/dl the night prior to the call and they passed out.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime test, excessive no delivery test and dat test at 0.08730 inches.No over delivery anomaly or under delivery anomaly noted.The stop current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and unexpected restart.Device uploaded properly using carelink.Device had minor scratched display window and cracked reservoir tube lip.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAH
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8517221
Report Number3004209178-2019-82113
Device Sequence Number1628928
Product Code OZO
UDI-Device Identifier00643169736078
UDI-Public(01)00643169736078
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-751NAH
Device Catalogue NumberMMT-751NAH
Device Lot NumberA6751NAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2019
Initial Date Received by Manufacturer 04/05/2019
Supplement Date Received by Manufacturer06/06/2019
Initial Report FDA Received Date04/16/2019
Supplement Report FDA Received Date06/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNOMED SET
Outcome Attributed to Adverse Event Hospitalization;
Patient Age48 YR
-
-