• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSET¿; INSET I 60/6 GREY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSET¿; INSET I 60/6 GREY Back to Search Results
Lot Number 5147653
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Elevated ketones/Diabetic Ketoacidosis (2364)
Type of Reportable Event Serious Injury
Event or Problem Description
Unomedical reference number (b)(4).Event occurred in the united states.The patient faced issued with the infusion set in the form of bent cannula, which was noticed after removing it.Prior to hospitalization, she self-treated with a bolus from the pump.Reportedly, the blood glucose was 1004 mg/dl and she experienced diabetic ketoacidosis, which led to hospitalization.She received intravenous insulin drip which resolved the issue.The infusion set was used for one day.She was admitted for two days.She stated that she was now on cardiac medicine after the hospitalization.Unomedical do not see bent/kinking as being related to human factors, but rather as a training issue including correct choices of insertion sites and infusion sets and cannula length.Furthermore, the soft cannula is a flexible material that during use and upon removal can bend slightly.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSET¿
Common Device Name
INSET I 60/6 GREY
MDR Report Key8517433
Report Number3003442380-2019-01245
Device Sequence Number16512469
Product Code FPA
UDI-Device Identifier05705244006874
UDI-Public05705244006874
Combination Product (Y/N)N
PMA/510(K) Number
K032854
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device No Information
Device Expiration Date06/01/2019
Device Lot Number5147653
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2017
Initial Report FDA Received Date04/16/2019
Patient Sequence Number1
-
-