MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRITANIUM REVISION ACETABULAR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRITANIUM REVISION ACETABULAR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 509-02-56E
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code Injury (2348)
Date of Event 03/19/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Device not returned.
 
Event or Problem Description
The patient presented with aseptic loosening of the acetabular cup, which was revised in (b)(6) 2018.The femoral stem was left in place during the first revision.The cup loosened again, requiring a second revision.When we exposed the components, we noticed that the mdm liner/acetabular shell wore a notch into the neck of the femoral stem.We removed all components and inserted a restoration modular stem, multihole acetabular cup and constrained liner.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRITANIUM REVISION ACETABULAR
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8518012
Report Number0002249697-2019-01617
Device Sequence Number1465802
Product Code LPH
UDI-Device Identifier07613327025156
UDI-Public07613327025156
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Catalogue Number509-02-56E
Device Lot Number41421M
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2019
Initial Report FDA Received Date04/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age63 YR
-
-