| Model Number |
MSERIES BI-PHASIC |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the device and the reported malfunction was not replicated or confirmed.The device was put through extensive testing without duplicating the malfunction.The devices control board to system board flex cable was replaced as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device powered on without display.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|