Model Number 5392
Device Problems Display or Visual Feedback Problem (1184); Operating System Becomes Nonfunctional (2996)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/11/2019
Event Type  Injury  
Manufacturer Narrative
Analysis confirmed the customer comment of a generated error and it was attributed to the main printed circuit board being out of specification in an electrical manner.All found defective parts were replaced and all other identified issues were resolved.The firmware was updated as was the rapid atrial pacing.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient was being 100% paced when the device abruptly stopped functioning.The clinician noted a "error 0810" message on the screen with instructions to contact the company.The device was taken out of service and replaced with a different temporary pacer.The generator was returned for service.The patient suffered no adverse outcome as a result of this event.It was noted that the generator had recently been received back from a calibration procedure with the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
lisa robertson
8200 coral sea st ne
mounds view, MN 55112
7635262723
MDR Report Key8518405
MDR Text Key142066293
Report Number3004593495-2019-00338
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00643169508767
UDI-Public00643169508767
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 04/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2019
Is the Reporter a Health Professional? No
Date Manufacturer Received04/03/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/19/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-