Model Number 4362022L
Medical Device Problem Codes Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/19/2019
Type of Reportable Event Serious Injury
Event or Problem Description
Post dilation, it was difficult to remove the savvy long balloon catheter via the sheath as it became elongated and detached from the catheter.There was no patient injury.The device was removed from the patient with use of a goose neck snare.The target lesion is the anterior tibial artery which was moderately calcified with a combination of stenoses and occlusions, approx 20 cm in length.The device was prep according to the instruction for use with no defects noted during prep.A constant and dedicated saline source was utilized through the balloon during prep.There were no difficulties removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product the product into the patient.A competitor's contrast media and inflation device was used with a ration of sixty/forty.The inflation device was successfully used with other devices.There was no resistance/friction while inserting the balloon through the rotating hemostatic valve.The catheter has never been in an acute bend.A 4fr brite tip sheath was used and then upsized to 6fr brite tip sheath.There were no difficulty inflating the balloon.The device is available for analysis as the customer has kept the balloon.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAVVY LONG 2.0X220MM L=120CM
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD
enniscorthy
co. wexford 0000
EI  0000
MDR Report Key8520326
Report Number1016427-2019-02729
Device Sequence Number1278133
Product Code LIT
UDI-Device Identifier00801741132742
UDI-Public00801741132742
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K072947
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Physician Assistant
Type of Report Initial
Report Date (Section B) 04/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2021
Device Model Number4362022L
Device Catalogue Number4362022L
Device Lot NumberCMCV0276
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date03/20/2019
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-