• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723LNAS
Medical Device Problem Code Device Difficult to Program or Calibrate (1496)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Insulin pump received with intermittent button response due to corroded keypad traces.No button error alarm during testing.No unlocked j2/lcd keypad connector.Unit passed self test.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
 
Event or Problem Description
The customer's parent reported via phone call that their son insulin pump had keypad anomaly.The customer¿s blood glucose level was 206 mg/dl.The customer reported that the insulin pump was not working and all buttons were not working.The customer reported that the time clock was advancing.The customer was advised to discontinue use of the pump and revert to backup plan per health care professional instructions.The customer was advised the pump will need to be replaced.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key8521553
Report Number2032227-2019-03188
Device Sequence Number16740265
Product Code OYC
UDI-Device Identifier00613994904485
UDI-Public(01)00613994904485
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723LNAS
Device Catalogue NumberMMT-723LNAS
Device Lot NumberA000202771
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2019
Initial Report FDA Received Date04/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age20 YR
Patient Weight132
-
-