| Model Number |
AA61181002 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
After receiving this complaint, we searched for other complaint and we didn't find other complaint on the lot n° 6338361.The balloon burst issue is known and monitored on a monthly basis.Possible root cause is a weakness area in the silicone material of the balloon.No other corrective action will be implemented as the specific trend for balloon burst of this product family is considered acceptable as per the threshold.
|
| |
|
Event or Problem Description
|
|
According to the available information, during the test of the balloon with 15ml of water for injection, the balloon busts before insertion.Not used on patient.
|
| |
|