WRIGHTS LANE SYNTHES USA PRODUCTS LLC 11MM/130 DEG TI CANN TFNA 400MM/RIGHT - STERILE; ROD,FIXATION,INTRAMEDULLARY
|
Back to Search Results |
|
| Model Number |
04.037.160S |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
03/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Expiration date is unknown( the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.The investigation could not be completed; no conclusion could be drawn at the time of filing this report.A review of the device history records has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2019, during a trochanteric fixation nail advanced (tfna) procedure, the nurse handed the tfna nail to the scrub technician, the nail was attached to the connecting unknown screw and the unknown insertion handle.The scrub technician mistakenly used the 5mm flexible driver to connect the tfna nail to the insertion jig instead of correctly using the 8mm hex driver with a connecting screw.The tfna nail was inserted.During the helical blade insertion, a portion of the tfna system, the blade became lodged halfway into the nail.After multiple extraction techniques, the blade was successfully removed and the nail followed.After examining the implants and seeing the damage, the doctor asked for a new tfna nail and tfna blade.There were 20 minutes of surgical delay.Procedure ended successfully.Patient outcome was successful.Concomitant device reported: unknown screws (part# unknown; lot# unknown; quantity 1) unknown nail insertion handles (part# unknown; lot# unknown; quantity 1) unknown flexible driver (part# unknown; lot# unknown; quantity 1) this complaint involves two (2) devices.This report is for one (1) 11mm/130 deg ti cann tfna 400mm/right - sterile this report is 2 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: d4: expiration date.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10: additional narrative: h3, h4, h6: part 04.037.160s, lot h679566: manufacturing date: july 16, 2018.Manufactured by synthes monument.Expiration date: may 31, 2018.No non-conformance reports (ncr's) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.H3, h6: a product investigation was completed: upon visual inspection, the device was found to be having damaged internal threads due to some unintended forces during the procedure near the locking mechanism.The remaining portions of the device show no other damage.The received condition of the device agrees with the complaint description.Functional testing of the received device is performed by inserting the helical blade into tfna nail.The helical blade was sliding into the tfna nail as it was intended to be.Functional test was performed to rotate the locking screw using the appropriate mating instrument.The screw was rotating but the locking and unlocking functions were not working due to the damaged internal threads of the tfna nail near locking mechanism.Dimensional inspection indicated the relevant dimension was within specification.The relevant tabulated drawing was reviewed during investigation; no design issues or discrepancies were observed during the investigation.A visual inspection and document/specification review were performed as part of this investigation.And functional test was performed according to the instructions but could not replicate the complaint condition.The complaint condition was confirmed, as the internal threads of the device was found to be damaged due to the unintended forces applied to remove the helical blade from tfna nail.While a definitive root cause cannot be determined, it is possible that the helical blade was removed from tfna nail while the lock screw was deployed.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition.Based upon these findings, no additional corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|